Working & Features of Vertical Laminar
Flow Unit
The Horizontal Laminar airflow provides
product proteciton within the work area,
which is confined by clear glass panels
and sliding glass doors.
Room air
is taken in from the side of the cabinet
through a pre-filter with 90% efficiency
which traps particles of 10µ
and above.
Air is spread evenly across the HEPA
filter, thereby providing a steam
of clean air withing the work area
of the cabinet; this removes all contaminated
air from the interior.
Perforated
Stainless Stell diffuser screen protects
the HEPA filter and maintain uniform
air flow velocity within the work area
Vertical
Laminar Flow occupies less space and
can be installed in a small area.
Provides
extra workspace (Depth of 29.5
/ 750mm)
Advantage
of working from both the sides and visibility
from all sides.